* 16-bit and 32-bit versions return substantially different (20-50%) magnetic field response to both Y and Z fields in Z-mode, although the sensor conditions are the same. 32-bit version reports increased gain for the Y signals, and decreased gain for the X signals
* The op-amp phase response rolls for 32-bit version for about 30 degrees in 0-200 Hz; the phase is flat in the 16-bit version. The phase behavior could be caused by a 0.5ms shift between the reference data and the input data.
* The op-amp response: <s>phase at zero frequency is flipped by 180 degrees in 32 bit version</s> - fixed
* Calibration chirps are cropped too much in the beginning
* Sometimes the subsequent calibration chirps come out at different DC levels. The magnetometer output looks like there is a sudden field jump between the chirps. Sometimes the jumps are larger than the magnetometer range. The jumps seem to be bigger when the field is not completely zeroed. '''Both 16-bit and 32-bit versions do that.'''
|