* 16-bit and 32-bit versions return substantially different (20-50%) magnetic field response to both Y and Z fields in Z-mode, although the sensor conditions are the same. 32-bit version reports increased gain for the Y signals, and decreased gain for the X signals
* The op-amp phase response rolls for 32-bit version for about
* Calibration chirps are cropped too much in the beginning
* Sometimes the subsequent calibration chirps come out at different DC levels. The magnetometer output looks like there is a sudden field jump between the chirps. Sometimes the jumps are larger than the magnetometer range. The jumps seem to be bigger when the field is not completely zeroed. '''Both 16-bit and 32-bit versions do that.'''
|