Optical Cryostat (Janis ST-100)
Jump to navigation
Jump to search
Files
There is a hard copy of the manual in the lab.
Notes
Here are some random useful pieces of information to know gathered mostly from emails with Dan Logan from Janis:
- The ST-100 is designed to hold and operate at Torr and the seals should hold this pressure for "several hours if not several days".
- It seems like this is not the case.
- With continuous evacuation with a turbo pump the pressure should be able to reach or Torr, the ladish gaskets on the cryostat aren't able to hold this high of a vacuum without the pump on though.
- The performance of the cryostat will be better but the pressures as low as Torr can be acceptable.
- The ST-100 should not be baked out because the window block and electrical feedthroughs are sealed using epoxy and baking out the cryostat might cause the epoxy to melt.
- What we've seen is that at room temperature while continuously pumping with the Pfeiffer turbo pump the pressure is pretty constant around 5.6 mbar (~4.2 Torr). While the cryostat is held around 5 K and continuously pumped down, the pressure gets as low as 4 mbar (~3 Torr).