Lena/Aug 2016

From zhivun
Revision as of 17:11, 22 August 2016 by ElenaZhivun (talk | contribs) (Created page with "= August 2016 = == 08/15/2016 == Talked to Dima about the NURI magnetometer network paper. His comments: * The device section describes a solution without a problem. Add to...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

August 2016

08/15/2016

Talked to Dima about the NURI magnetometer network paper. His comments:

  • The device section describes a solution without a problem. Add to the introducion:
 * need for the accurate timing
 * need to correlate signals from the different sensors
 * magnetometer has to be portable

Talked to Thad about a possibility to have a synchronously pumped scalar Mx magnetometer in the fMCG project. The advantages would be:

 * not having to cancel the field
 * no calibration
 * no 1/f electronic noise

Disadvantages:

 * spin-exchange relaxation
 * no vector sensitivity