* '''[FIXED]''' The op-amp response: phase at zero frequency is flipped by 180 degrees in 32 bit version
* '''[FIXED/not reproduced?]''' Calibration chirps are cropped too much in the beginning
* '''[FIXED in 32-bit version]''' Sometimes the subsequent calibration chirps come out at different DC levels. The magnetometer output looks like there is a sudden field jump between the chirps. Sometimes the jumps are larger than the magnetometer range. The jumps seem to be bigger when the field is not completely zeroed. '''Both 16-bit and 32-bit versions do that.'''
* '''[FIXED]''' Calibration chirps go too far out in the frequency domain and reach 1kHz, the modulation frequency. This causes the X field to not be demodulated incorrectly. '''Both 16-bit and 32-bit versions do that.'''
 