Lena/Dec 2016: Difference between revisions

Jump to navigation Jump to search
(Created page with "= December 2016 = == BACK/NEXT == << BACK NEXT >>")
 
 
[[Lena/Nov 2016|<< BACK]] [[Lena/Jan 2017|NEXT >>]]
 
== 12/07/2016 ==
 
The signal amplitude from the OTS cell seems to be smaller than from the buffer gas cells. I talked to Mike and we are concerned that this might be because of the large field gradients due to the small shell coils. It might be possible that the field is not zero in all parts of the cell.
 
For this reason I've made a larger Helmholtz coil system. Here is the calibration of the coil system.
There is also an anti-helmholtz coil in the Y direction, but it's not connected and not calibrated.
 
=== Calibration data ===
 
{| class="wikitable"
|-
! Direction !! B at Imax (240 mA), nT !! B at -Imax (-240 mA), nT
|-
| X || 11380 || -10250
|-
| Y || 13180 || -11600
|-
| Z || 8140 || -9760
|}
 
=== Coil calibration ===
 
{| class="wikitable"
|-
! Direction !! nT/mA
|-
| X || 45.06
|-
| Y || 51.95
|-
| Z || 37.29
|}
 
=== Shell pinout (left to right) ===
 
# orange, yellow = X coil
# green, blue = Y coil
# purple, gray = Z coil
# black, white = thermoresistor

Navigation menu