Lena/Oct 2016: Difference between revisions

Jump to navigation Jump to search
 
== 10/26/2016 ==
 
=== Pregnant patient in WIMR ===
 
Measured a pregnant patient in WIMR. Although all channels worked properly and probe sensitivity was 1 fT/sqrt(Hz), we didn't get any good data, and couldn't see the fMCG signal at all. Maybe there were large field gradients that we didn't properly get rid of, like from wires in a bra or other clothes. The patient returns on November, 14th.