'''8/10/14'''
Running more tests with the PDA36A. I checked it from 5 to 100 nW as measured on the power meter. (again though,Using I'mthe notcone veryhelped confidentcut in it's measurements at the nW level--lots ofdown fluctuations anddown harda tolot. get 0 offset) andI found the response to be logarithmic when the gain was set to 0 dB as how I was using it on 8/8/14. I don't think this will work unless I can more accurately measure nW level beams with the power meter and then make a good regression equation for a logarithmic response. I then repeated these measurements with the gain at 60 dB and found the response to be much more linear. Data is [https://wiki.physics.wisc.edu/yavuz/images/2/29/Thorlabs_pda36a_response.xlsx here]. Maybe I can find the best gain setting that gives the most linear response at these powers and get a good linear fit at slightly higher beam powers and extrapolate to the single nW level even though I can't measure those powers well with the power meter.
 
I tried a few other settings and 60 dB seems like a good balance.
 
 