Zjs log

From Yavuz Group
Revision as of 04:23, 7 May 2014 by Zjsimmons (talk | contribs) (Created page with "--~~~~ Figured out that the flat polished fibers were the cause of power fluctuations after the TA in 1055nm set-up. Using an angle polished fiber fixed the issue. Power fluc...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

--Zjsimmons (talk) 23:23, 6 May 2014 (CDT)

Figured out that the flat polished fibers were the cause of power fluctuations after the TA in 1055nm set-up. Using an angle polished fiber fixed the issue. Power fluctuations are also not the reason the circuit is not locking the SHG cavity. Perhaps the gain roll-off needs to be at lower freq as we are seeing some wiggles around 420Hz (piezo resonance freq).