Exfo Wavemeter

From Yavuz Group
Revision as of 16:44, 4 January 2016 by Nbrewer (talk | contribs)
Jump to navigation Jump to search
WA-1500.jpg

Files

Notes

  • It seems like a couple times a year the wave meter needs to be realigned or have the reference laser recalibrate. The description of each error is in the manual above. There are also notes from various people who have realigned/recalibrated it through the years.
  • In December 2015 error 10 kept occurring. The description of error 10 in the manual is "The laser stabilization circuit is indicating an over temperature condition. Turn off the WA- 1500/1000, wait 15 minutes and power up again. It the problem persists, contact EXFO Service."