PKR 251 Pressure Gauge: Difference between revisions

Jump to navigation Jump to search
no edit summary
No edit summary
No edit summary
[[File:PKR251_Picture.pdfjpg|thumb|400px150px]]
 
==Files==
* [https://wiki.physics.wisc.edu/yavuz/images/e/ee/PKR251_manual_2008-04.pdf User Manual]
* [https://wiki.physics.wisc.edu/yavuz/index.php/File:PKR251_Cleaning_Rev_1.pdf Cleaning Instructions]
 
==Notes==
 
'''10/20/2014'''
 
* The pressure gauge is reading a constant 4e-3 mbar whether the cryostat is at room temperature or cooled down with liquid helium. This is too high of a pressure. Because we have not changed the anything with the cryostat that would have caused a leak in the vacuum and because the reading is not changing at all, I think the pressure gauge is not reading the correct thing rather than there being something wrong with something else in the system.
 
* I contacted Pfeiffer Vacuum and they sent the manual and cleaning instructions above. I took the gauge apart and cleaned it as instructed. I did not replace any parts.