==Notes==

Here are some random useful pieces of information to know gathered mostly from emails with Dan Logan from Janis:

* The ST-100 is designed to hold and operate at <math>10^{-4}</math> Torr and the seals should hold this pressure for "several hours if not several days".
::It seems like this is not the case.

* With continuous evacuation with a turbo pump the pressure should be able to reach <math>10^{-5}</math> or <math>10^{-6}</math> Torr, the ladish gaskets on the cryostat aren't able to hold this high of a vacuum without the pump on though.

* The performance of the cryostat will be better but the pressures as low as <math>10^{-3}</math> Torr can be acceptable.

* The ST-100 should not be baked out because the window block and electrical feedthroughs are sealed using epoxy and baking out the cryostat might cause the epoxy to melt.

* What we've seen is that at room temperature while continuously pumping with the [[Turbo Pump Station (HiCube 80 Eco)|Pfeiffer turbo pump]] the pressure is pretty constant around 5.6<math>*10^{-5}</math> mbar (~4.2<math>*10^{-5}</math> Torr). While the cryostat is held around 5 K and continuously pumped down, the pressure gets as low as 4<math>*10^{-8}</math> mbar (~3<math>*10^{-8}</math> Torr).

Latest revision as of 15:41, 22 July 2014

ST-100.JPG

Files

There is a hard copy of the manual in the lab.

Technical Drawing


Notes

Here are some random useful pieces of information to know gathered mostly from emails with Dan Logan from Janis:

  • The ST-100 is designed to hold and operate at [math]\displaystyle{ 10^{-4} }[/math] Torr and the seals should hold this pressure for "several hours if not several days".
It seems like this is not the case.
  • With continuous evacuation with a turbo pump the pressure should be able to reach [math]\displaystyle{ 10^{-5} }[/math] or [math]\displaystyle{ 10^{-6} }[/math] Torr, the ladish gaskets on the cryostat aren't able to hold this high of a vacuum without the pump on though.
  • The performance of the cryostat will be better but the pressures as low as [math]\displaystyle{ 10^{-3} }[/math] Torr can be acceptable.
  • The ST-100 should not be baked out because the window block and electrical feedthroughs are sealed using epoxy and baking out the cryostat might cause the epoxy to melt.
  • What we've seen is that at room temperature while continuously pumping with the Pfeiffer turbo pump the pressure is pretty constant around 5.6[math]\displaystyle{ *10^{-5} }[/math] mbar (~4.2[math]\displaystyle{ *10^{-5} }[/math] Torr). While the cryostat is held around 5 K and continuously pumped down, the pressure gets as low as 4[math]\displaystyle{ *10^{-8} }[/math] mbar (~3[math]\displaystyle{ *10^{-8} }[/math] Torr).